AMI-300 fully automatic temperature-programmed chemical adsorption instrument
Product Details
Core Specifications
Applications
Detailed Description
AMI-300 series fully automatic temperature-programmed chemical adsorption instrument
AMI-300 chemical adsorption analyzer is a new generation of fully automatic temperature-programmed chemical adsorption instrument that can perform dynamic temperature-programmed catalyst characterization experiments (TPR, TPO, TPD, pulse chemical adsorption, etc.); it can measure metal dispersion, relative activity, and adsorption intensity, and the test time is only 1/3 of the traditional volumetric method. Depending on your needs, the standard TCD detector can be used for gas analysis, or it can be combined with a mass spectrometer or other detectors (FID, FTIR, GC, etc.).

Features:
Multiple air intakes Four high-precision mass flow controllers (MFC) are optional, expanding to 12 air inlets.
temperature range The experimental temperature can reach 1200℃, the heating rate in the whole range can reach 1-50℃/min, and the optional low-temperature component temperature can be as low as -130℃
Peak spread is small 1/16-inch 316 stainless steel pipeline ensures a small dead volume, which is beneficial to improve the signal response rate and reduce the spread of peaks.
Built-in saturated steam generator Used to generate gas with saturated liquid vapor, the temperature of the saturated steam generator is controllable.
Easy sample loading and unloading Flexible and movable shell-type heating furnace with various specifications of sample tubes, suitable for different sample sizes and dosages to meet user testing needs
Short analysis time The automatically controlled air cooling component enables faster cooling and effectively shortens the experiment time.
Multi-channel temperature control The temperature of the heating furnace or sample bed can be freely switched to control the heating rate during instrument testing, and the temperatures of the heating furnace and sample bed can be recorded in real time for data analysis.
High safety factor It provides multi-directional temperature detection, over-temperature protection system, TCD flow monitoring anti-dry burning system, front emergency switch and other options to provide better safety options.
Measure accurately Equipped with a 4-filament high-precision thermal conductivity cell detector (TCD) and quantitative loops (Loop) with different ranges. Automatic or manual quantification can be selected to maximize sensitivity and compatibility.
Flexible user interface Based on Windows operating system software, the program sets up the experimental process, controls instrument functions and data processing.
Own operation, fully automatic testing The experiment is run fully automatically, and the computer automatically collects and stores data. In the advanced user mode, the instrument setting window is fully open, allowing users to operate with a high degree of freedom.
Can be connected to a mass spectrometer (MS) or gas detector Supports multiple external detectors, provides series & parallel connection methods, and can embed mass spectrometry (MS) data collection into AMI software to export TCD & MS data in the same file
No vapor condensation and adsorption retention All pipelines and valves in the instrument can be temperature controlled to prevent condensation and adsorption retention of steam.
No need to purchase additional gas mixer Built-in gas mixer can provide arbitrary and uniform gas mixture. This gas mixer is also suitable for fully automatic multi-point BET specific surface area analysis
Chemical versatility and high sensitivity Appropriate sealing materials and TCD filaments can be selected according to different experimental needs.
Performance parameters:
modelAMI-300
Typical sample0.1-1.0g
temperature rangeRoom temperature°C——1200°C
Low temperature option-130°C——1200°C
heating rate1°C/min——50°C/min
Standard operating pressure (TPX units)atmospheric pressure
Optional pressure range (high pressure)30bar/100bar
Gas input port (low pressure)4 channels of carrier gas, 4 channels of process gas, 2 channels of mixed gas
Gas flow rate5——50cm3/min (standard mode)
Sample tube type (low pressure)Quartz U-shaped tube, bubble tube, straight wall tube
Sample tube type (high pressure)316 stainless steel reaction tube
TCD detectorTwo materials available (tungsten; gold/tungsten)
Pipe material316 stainless steel, 1/6 inch
sealing ringOptional Viton, Buna-N, Kalrez, etc.
sizeWidth 56cm; height 60cm; depth 61cm
weight55Kg
power supply220—240V,50/60Hz
Typical applications:
Studying the surface active sites and quantity, strength, activity, stability, selectivity and deactivation of catalysts is very important for industrial reaction processes. Surface activity plays a vital role in the development of materials for the catalytic, chemical and petrochemical industries, such as fine chemicals, fuels, fertilizers, emission controllers, batteries, fuel cells and energy storage. Heterogeneous catalysts are also widely used in catalytic cracking and reforming reactions, hydrogenation reactions (hydrodesulfurization, hydrodenitrification, hydrodeoxygenation, hydrodemetallization), selective oxidation and reduction reactions, automobile exhaust pollution control, hydrocarbon isomerization, Fischer-Tropsch process, water-gas shift and many other important industrial reactions.
Product Overview
The AMI-300 series fully automatic temperature-programmed chemical adsorption instrument is a multi-functional catalyst characterization equipment that can perform experiments such as TPR, TPD, TPO and pulse chemical adsorption, and is used to determine metal dispersion, relative activity and adsorption strength. It features multiple air intakes, flexible temperature control, and rapid cooling, and can be used in conjunction with a mass spectrometer or gas detector to ensure high-precision measurement and efficient experimental processes. Suitable for catalyst research and development in catalysis, chemical and petrochemical industries.