Don’t worry about controlling your sugar levels in summer! Shenghan's self-developed sugar column: stable performance, green and low carbon, perfectly adapted to GB 5009.8-2023!

Summer happiness comes from ice cream, iced drinks and desserts;
But healthy peace of mind isAccurate sugar measurement and scientific sugar controlGiven.
As the temperature rises, ice cream, sugary drinks, and desserts are ushering in the peak consumption season. Food companies, third-party testing, and food and drug supervision have a surge in demand for quantitative testing of glucose, fructose, sucrose, lactose, and maltose.
GB 5009.8-2023 "National Food Safety Standard Determination of Fructose, Glucose, Sucrose, Maltose and Lactose in Foods" clearly requiresAccurate, stable and repeatableFor sugar analysis solutions, ion chromatography-pulsed amperometric method is the mainstream choice.

Recently, many laboratories have encounteredimportsugar columnTight supply, extended delivery times, and rising costsThe problem is that the detection progress is frequently stuck.
Sheng Han self-developedSH-CAC-2A sugar chromatography column, stable performance,Domestic spot goods, green and low carbon, supply chain security,It meets the national standard requirements of GB 5009.8-2023 and has submitted a full score answer sheet with complete verification data.
01
Shenghan’s self-developed sugar column application solutions
Sample preparation
Take 2g of solid sample and 10g of semi-solid or liquid sample in a 100mL colorimetric tube, and add about 50mL of water.Vortex to mix, then add 2 mL of acetic acid solution and mix well, sonicate for 30 min, transfer to a 100 mL volumetric flask and dilute to volume with water, mix well, and let stand for 20 min. After the sample extraction solution is diluted to an appropriate multiple, it is passed through a C18 solid-phase extraction cartridge and a 0.22 μm filter membrane in sequence. The remaining filtrate is taken for detection. It is qualitatively determined based on the retention time and passed through the chromatograph.Peak areaor peak height quantification.
Chromatographic conditions
Device model | CIC-D200E |
column | SH-CAC-2A |
flow rate | 1.0mL/min |
Eluent conditions | A: Pure water; B: 200 mMNaOH solution 0.0 -16 min 80%(A)+20%(B) 16.1 -30 min 0%(A)+100%(B) 30.1-40 min 80%(A)+20%(B) |
Column temperature | 35℃ |
pressure | 13MPa |
quantitative loop | 10μL |
Potential waveform diagram |
|
Detection method | pulsed ampere mode |
working electrode | Flexcell gold electrode |
Sample spectrum





▲5Mixed standards (glucose, fructose, sucrose, lactose, maltose) linear

▲Sample spectrum
After testing, the self-developed sugar chromatography column SH-CAC-2A has good separation effect. The linear correlation coefficients of glucose, fructose, sucrose, lactose and maltose are all above 0.999, which can meet the linear range requirements of 2.5-25mg/L specified in GB 5009.8-2023. When tested on real samples, the column achieved normal separations. This shows that this column can meet the detection requirements of GB 5009.8-2023 for five sugars in food.
02
Advantages of Shenghan's self-developed sugar column

▷Wide application: Easily meet the national standard GB 5009.8-2023 and various sugar analysis needs in other fields, one column can be used for multiple purposes.
▷Strong performance: high column efficiency and high resolution, bringing accurate and reliable analysis results.
▷Long life: Excellent physical and chemical stability, 1000 needles are still stable in continuous operation, reducing use costs.
▷Substitutability: It can perfectly replace similar products of imported brands, and the analysis results are highly consistent, ensuring supply chain security and cost optimization.
Green low-carbon laboratory selects Shenghan’s self-developed sugar columnNot only a performance column, but also a low-carbon column:
○Domestic and self-developed, shortening the supply chain and reducing transportation carbon footprint;
○Adapted to pure water/low organic elution system to reduce organic solvent consumption;
○High column efficiency and long life, reducing the frequency of consumable replacement and solid waste;
○Full-process green production to support laboratoriesESGwith the dual carbon target;
○Under the twin trends of sugar control and low carbon, it is not only a powerful detection tool, but also responsible for environmental protection.
03
Summary: If you want to test your sugar in summer, this is the right product to choose.
Faced with the shortage of imported sugar columns, long delivery times and price fluctuations, Shenghan self-developed sugar columns provideStable spot + fast delivery + full technical support. Precisely controlling the sugar content with professional strength, adhering to the green and low-carbon original intention to practice the concept of environmental protection, and adhering to the original intention of quality with domestic ingenuity, Shenghan Chromatography makes summer sweetness both healthy and fresh green, so that every test is safe and reliable, and you can calmly go to the refreshing summer time.