Tianlong Technology fully automatic nucleic acid extraction instrument GeneRotex
Product Details
Core Specifications
Applications
Detailed Description
Fully automatic nucleic acid extraction instrument
Tianlong Technology's fully automatic nucleic acid extractor GeneRotex uses independently developed Rotary Mixing Technology (RMT), which performs rapid 360° rotation and mixing based on slow up and down reciprocating motion. This technology has applied for an invention patent (Patent No. ZL201710435931.8), which can control the risk of cross-contamination caused by nucleic acid aerosols to a lower level and run more quietly. In addition, design concepts such as "independent modules" and "one machine, three models" are also fully reflected in this series, which can meet the diverse downstream experimental testing needs.
- Rotary nucleic acid extraction technology
- Extraction modules 24, 48, 96 are optional
- Various measures to control pollution
- Ultra-quiet operation
01
Various extraction modules
96 throughput: The innovative 6×16 extraction module, equipped with a self-developed 96 deep-well plate and a single six-strip tube, ensures high-throughput extraction while overcoming the shortcomings of inflexible throughput and waste of reagents in traditional extraction modules. From now on, you can choose from 1 to 96.
48-throughput: A large-system extraction solution with a sample volume of 1 mL, combined with self-developed consumables, makes ultra-sensitive detection of infectious pathogens as low as 2 copies/mL possible.
24-throughput: Using an innovative 5mL ultra-large sample volume and a maximum processing volume of 15mL, it can meet the requirements of various testing projects such as prenatal DNA testing and tumor targeted drug guidance.
02
Built-in 7-inch full-color LCD screen, touch operation, or one-click operation by scanning the QR code.
Rotating mixing: In order to minimize the generation of aerosols, this equipment adopts independently developed RMT technology. Compared with traditional oscillating mixing technology, the generation of aerosols is greatly reduced. The risk of false positives caused by cross-contamination is more effectively controlled, further ensuring the accuracy of experimental results.
Negative pressure HEPA: The instrument has a built-in replaceable high-efficiency air filter and a negative pressure exhaust device, which greatly reduces the risk of pollution from exhaust gases.
UV disinfection: The instrument has a built-in UV lamp and can flexibly set the disinfection time.
Automatic hatch: An independent motor-driven component eliminates the need to manually open and close the hatch, reducing operator contact with potential biological hazards and reserving space for automatic sampling equipment.
04
As the fifth generation of Tianlong Technology’s fully automatic nucleic acid extraction instrument, GeneRotex 96 inherits the classic 96-well deep well plate and innovative single six-strip tube, and is compatible with all types and specifications of Tianlong’s extraction kits. In addition, this series of products is equipped with special large-system nucleic acid extraction reagents, which can greatly improve work efficiency and detection sensitivity.
05
High extraction efficiency
The nucleic acid recovery rate is >95%, the detection rates of 10lU/mLHBV and 30lU/mLHCV are both ≥98%, and it can even extract 2 copies/mL of COVID-19 standard.
| Instrument model | GeneRotex 96 | GeneRotex 48 | GeneRotex 24 |
| Medical device registration certificate | Shaanxi Machinery No. 20150054 | Shaanxi Machinery No. 20150054 | Shaanxi Machinery No. 20150054 |
| sample throughput | 1-96 | 1-48 | 1-24 |
| Processing volume | 30-1000μL | 50-3000μL | 50-15000μL |
| Recommended sample size | 200μL | 1000μL | 5000μL |
| Applicable consumables | 96-well deep well plate, Customized single six-unit pipe | Customized 48-well deep well plate | Customized ultra-large volume single seven-piece pipe |
| Rotation speed | ≤3000rpm | ||
| Temperature control range | Cracking heating: room temperature ~ 120℃ Elution heating: room temperature ~ 120℃ | ||
| Mixing method | Swirl to mix | ||
| operating language | Built-in Chinese/English operating languages, which can be switched freely | ||
| Operation mode | 7-inch full-color LCD touch screen or code scanner control | ||
| program storage | Built-in mode program, can store up to 10,000 programs | ||
| Program management | Flexible to create, edit, apply and delete programs | ||
| automatic hatch | Motor driven automatic switch experimental cabin | ||
| QR code scanning | Can be connected to an external scanner, automatically recognizes applications after scanning, and runs them with one click | ||
| Power outage protection | You can freely choose whether to continue running the experiment when the power is unexpectedly cut off and then restored. | ||
| pollution control | Negative pressure HEPA exhaust filter module with built-in UV disinfection module | ||
| Instrument interface | USB interface | ||
| weight | 45kg | ||
| Overall dimensions | 490mm (L)×510mm (W) ×480mm (H) | 490mm(L) x540mm(W) X480mm(H) | |
| Supply voltage | AC 220V,50Hz | ||
| temperature range | 10℃~30℃ | ||
| relative humidity | ≤80% | ||
Product Overview
The GeneRotex series of fully automatic nucleic acid extractors adopt self-developed rotary mixing technology, which can effectively reduce the risk of cross-contamination caused by aerosols and operate more quietly. The equipment provides 96, 48, and 24 flux modules, with flexible operation methods, efficient contamination control and good reagent compatibility, and is suitable for diverse downstream experimental detection needs.
Send an inquiry
Recommended Products
Tianlong Technology Nucleic Acid Extraction GeneMix Pro Sample Processing System
US$ 41958.00
Model:GeneMix Pro