Fully automatic high-throughput specific surface area and pore size analyzer
Product Details
Core Specifications
Applications
Detailed Description
BSD-660 Series Physical Adsorption Instrument Selection Guide | |||||
Instrument model | |||||
Instrument name | |||||
Main functions | BET specific surface area analysis, Mesoporous pore volume and pore size analysis. | In addition to the 660S functions, With micropore volume and pore size analysis Structure and function. | In addition to the 660M function, Capable of adsorption analysis of various gases (non-corrosive) Structure and function. | In addition to the 660MG function, Capable of corrosive gas adsorption analysis Knotstructure and function. | In addition to the 660MG function, With vapor adsorption analysis Structure and function. |
Commonly tested materials | Non-porous materials, such as lithium battery materials, Inorganic salts, oxides, Hydroxide, metal powder, organic materials, etc. | Porous materials, such as zeolite molecular sieves, activated carbon,MOF、COF、HOF, porous carbon, silica gel, etc. | |||
Main functions:
◆ Full function: specific surface area, mesopore, micropore, ultramicropore analysis;
◆ High-throughput: The number of analysis bits is optional: 3/6/9/12;
◆ Fully automatic: degassing → test, fully automatic switching;
◆ High configuration: Optional dual-station dual molecular pump set, molecular pump degassing;
◆ Anti-pollution: "Pressure-controlled heating" of microporous samples prevents pollution and degassing;
◆ Zero helium pollution: Helium temperature measurement area first, then automatic degassing;
◆ Automatic cycle: automatic evaluation of material cycle adsorption performance;
◆ Conventional gas adsorption: such as N2, O2, Ar, CO, CO2;
◆ Combustible gas adsorption: such as H2, CH4, C2H6 and other alkanes and alkynes;
Optional Function Optional Function
+ constant temperature water bath
Achieve non-corrosive gas adsorption at any temperature between -10-80℃
+Liquid nitrogen level constant device LNT
Realize non-corrosive gas adsorption measurement at any temperature from 100K to room temperature
Technical parameters:
◆ Wide testing range: specific surface area 0.0005㎡/g or more, pore diameter 0.35-500nm;
◆ Testing efficiency: multi-point BET (excluding degassing process), standard mode 12 samples/60min; extreme speed test mode 12 samples/15min;
◆ High testing accuracy: specific surface area, pore diameter, pore volume, adsorption capacity, quantitative error <0.5% RSD (based on BET value of standard sample);
◆ Programmed temperature rise and degassing: software controls program temperature rise, room temperature -400℃, accuracy better than 0.1℃;
◆ Intelligent degassing completion judgment: supports automatic software judgment and automatically judges the degassing effect according to pressure changes;
◆ Prevent flying and degassing: "Programmed pressure control" + "Programmed temperature control" + "Degassing furnace lift" = "Pressure controlled temperature rise";
◆ Vacuum degree: the vacuum degree can reach 10-9Pa; try
Feature structure:

Technical advantages:
◆ High throughput and high efficiency: supports the analysis of up to 12 samples at a time;
◆ True full automation: The world's first degassing furnace and Dewar cup automatically switch, no need to manually transfer sample tubes or degassing furnace;
Patent name: Fully automatic physical adsorption instrument that automatically switches the positions of the heating furnace and constant temperature bath cup. Patent number: ZL202020232044.8;
◆ High time utilization: It solves the waste of time when conventional instruments cannot start testing after degassing is completed after get off work, making it a reality to install samples after get off work and read data at work;
◆ Completely eliminate "helium pollution": Helium test dead volume → vacuum heating degassing → adsorption test. It is the first in the world to solve the problem of helium pollution in micropore analysis and improve test accuracy;
◆ Anti-flying and degassing: Supports "programmed pressure control" + "programmed temperature control" degassing, and automatically raises and lowers the degassing furnace according to pressure changes to prevent samples from flying;
Patent name: Physical adsorption instrument with programmable pressure-controlled anti-flying and degassing system Patent number: ZL202020230457.2;
◆ Support automatic cycle test: automatic degassing + test cycle test, used to evaluate material adsorption performance stability and adsorption performance lifespan evaluation;
"Pressure controlled temperature rise" anti-flying and degassing technology
That is, the programmed temperature rise technology controlled by gas pressure automatically starts and stops the program temperature rise according to pressure changes, fundamentally preventing the sample from flying;
Patent name: Physical adsorption instrument with programmable pressure control and anti-flying degassing system Patent number: ZL202020230457.2

"Automatic constant temperature zone" patented technology
The curve shows that the change rates of the "equivalent volume of the adsorption chamber" in the 24-hour temperature zone constant effect of the three methods are: BSD temperature zone servo automatic constant: 0.10%, isothermal clamp: 0.25%, liquid level sensor: 0.55%

Patent name: Physical adsorption instrument with automatic constant structure in temperature zone Patent number: ZL2018 2 0401132.9
◆ Convenient installation and sealing: One-time sealing technology for 6 sample tubes in a single analysis station, no need to seal each tube one by one, unparalleled efficiency experience;
Patent name: A physical adsorption instrument with dense multi-sample tubes and sealed test tube jackets Patent number: ZL 2019 2 1078195.6;
◆ The gas circuit system is fully constant temperature: the internal gas circuit system of the instrument is fully constant temperature to 40℃, with an accuracy better than ±0.01℃;
◆ Sliding door: humanized and easy to open and close, saving laboratory space;
Patent name: Physical adsorption instrument with upper and lower opening and closing protective covers Patent number: ZL202022203243.9;
◆ Electric turbine liquid nitrogen pump: humanized liquid nitrogen addition, stepless speed regulation, free movement, safe and convenient, liquid nitrogen pollution-free;
Patent name: An electric liquid nitrogen pump with impeller structure (non-pneumatic) Patent number: ZL 2017 2 0864873.6;
◆ High reliability: International supplier system, core components are imported with original packaging;
◆ Open data interface: The software can be used in conjunction with Lims to upload test results to Lims;
Comparison test:

Test report:


Awards received:
1.3i Award - Outstanding New Products in the Scientific Instrument Industry in 2022
2. Won the fifth "Good domestic instruments"Awards
Related information:
Product Overview
The instrument has full-functional specific surface area and pore size analysis capabilities, supports micropore, mesopore and ultra-micropore analysis, and provides 3/6/9/12-bit analysis position selection to achieve fully automatic degassing and testing switching. Its high configuration and anti-pollution design ensure test accuracy and safety. The unique "pressure controlled temperature rise" technology and "programmed pressure control and anti-flying" system significantly improve test efficiency and accuracy. In addition, the equipment supports high-throughput analysis, improves time utilization, and can automatically evaluate the cycle adsorption performance of materials.