Chloride ion meter is an electrochemical analysis instrument used to measure the concentration of ions in a solution. It is especially suitable for quality monitoring of high-purity water in power plants (such as steam, water, boiler feed water, etc.), and can also be used in the structures of colleges and universities.

The chloride ion meter instrument adopts a two-point calibration method. In order to meet the needs of various pNa value measurements, a set of calibration solutions with different pNa values ​​are used. Users can choose according to the pCL value measurement range. As a precision testing instrument, ensuring accuracy is the most basic requirement. So how to ensure the accuracy of chloride ion meter measurements?

1. Calibration is required every time the computer is turned on. It is recommended to operate as follows (power on, select "Calibration", put the electrode into the selected calibration solution (such as 0.00pCL), press the "△▽" key, and adjust the pNa value to 0. 00pCL Press the "Confirm" key to set the temperature, press the "△▽" key to set the temperature to the temperature value of the calibration solution, and then press the "Confirm" key. At this time, the instrument calibration point is calibrated and press the "Confirm" key to enter the second calibration point);

2. After measuring one sample and then measuring another, be sure to clean it, because when measuring a high concentration and a low concentration, the MV value will change, otherwise the samples will contaminate each other, so just clean it a little.

3. When testing a sample, the main thing is to see whether the MV value is stable or not. Once it is stable, it can be read.

4. The PCL value is displayed when the machine is turned on. Press the switch button and it will switch to the MV interface in about two or three seconds. Press the switch button again and it will switch to the concentration unit. The unit mg/L or microgram/L will be automatically selected according to the concentration.

5. When the instrument is measuring, if it is found that the display reading overflows, the measuring part is open circuit or the potential is too high, exceeding the specified value of 2000.0mV, the measuring part should be shut down to check whether there are bubbles, disconnections, etc.

6. When measuring pX, the positioning regulator cannot adjust to the pX value specified in the solution, which means that the zero potential of the measuring electrode is too different. At this time, the positioning compensator and slope compensator must be adjusted to offset this potential before measurement.