Gangyan Nanok ONH5500 Oxygen, Nitrogen and Hydrogen Analyzer
Core Specifications
Product Details
Product introduction
GangyanakAtIn 1977, it took the lead in developing a vacuum melted gas analyzer. In 1991, the pulse infrared oxygen analyzer was industrialized here, breaking the foreign monopoly. Through 40 years of technical accumulation, Gangyan Nano Oxygen Nitrogen and Hydrogen Analyzer has received good market feedback and customer recognition.
ONH5500 oxygen, nitrogen and hydrogen analyzer is a new oxygen, nitrogen and hydrogen analyzer launched by Gangyan Nanoco in 2020. It has outstanding features such as three-element joint determination, high-efficiency small-flow gas circuit, and intelligent software system. It can quickly and accurately measure the full range of oxygen, nitrogen, and hydrogen elements in solid inorganic materials such as steel, metal powders, nonferrous metals, ceramics, and minerals.
Technical parameters and advantages
Analysis scope | Hypoxia:0.1ppm~0.5% Nitrox: 0.5%~20% Low nitrogen:0.1ppm~0.5% High nitrogen: 0.5%~50% hydrogen:0.1ppm~2000ppm Changing the sample size can extend the analysis range to100% |
Analysis accuracy | nitrogen:1ppm or RSD≤1.0% hydrogen:0.2ppm or RSD≤2.0% Note: The standard deviation or uncertainty shall be no greater than the reference material. |
Sensitivity | 0.01ppm |
Analysis time | Generally3 minutes |
Sample weighing | Generally1g, the weighing amount can be changed according to the sample content |
pulsefurnace | current0~1500A, power 8.5KVA, maximum temperature up to 3500℃ |
carrier gas | High purity helium, purity≥99.9995% |
Power Qi | Ordinary nitrogen |
power supply | Pulse furnace:220VAC±10%,50±1Hz, current 50A Instruments and others:220VAC±10%,50±1Hz, current 10A |
weight | 140kg |
size | 630mm*760mm*760mm (width*depth*height) |
Instrument features
1. Using non-dispersive infrared technology(NDIR)Measure oxygen and hydrogen;Measured using highly sensitive thermal conductivity technologynitrogen, a single analysis can measure simultaneouslyO/N/H three elements
2. Efficient and new pulse heating furnace design can achieveHigh temperature above 3500℃, melting all kinds of solid inorganic samples
3. High efficiency and simple analysis gas path, saving expensiveHe carrier gas to reduce analysis costs
4. Humanized software operation mode, real-time monitoring of key parameters of the instrument;Enables software-controlled automatic leak detection and other rich self-diagnostic functions
5. Flexible software method settings, first-class linearization algorithm
6. Automatic switching between high and low range channels can be realized during the analysis process.
7. High stability flow control system to ensure the accuracy of analysis results
8. Modular detection unit design, free to choose and matchO/N/H element, efficient and reasonable