MDS-15 closed high-throughput microwave digestion/extraction/synthesis workstation
Core Specifications
Product Details
MDS-15 high-throughput sealed microwave digestion/extraction/synthesis workstation, a product on the list of "Domestic Good Instruments" in 2014, is built with Shanghai Xinyi's nearly 30 years of industry experience and R&D technology. It is multi-purpose, convenient and efficient, durable, safe and reliable, and can meet the needs of different users. It can also completely digest special samples such as petrochemical samples, polymer materials, animal and vegetable oils and so on. In addition, safe pressure relief technology, aerospace composite fiber outer tanks whose performance far exceeds that of PEEK, and PFA coatings with excellent anti-corrosion properties, all reflect Shanghai Xinyi's higher pursuit of instrument performance and product quality.
Product comprehensive technical characteristics

Sturdy and durable industrial-grade furnace cavity + safe and explosion-proof sliding furnace door to ensure the operator's personal safety
The wall thickness of the sturdy and anti-corrosion 316L stainless steel furnace cavity is more than 3mm, which can effectively absorb impacts and protect personnel safety under special circumstances. Double-sided multi-layer PFA spray coating on the inside and outside prevents acid mist from corroding the furnace cavity, extending the service life of the furnace cavity, and has a 5-year warranty commitment. The pop-up buffer safety explosion-proof furnace door makes operation more reliable and safer.

The Safety Bolt design of the safety pressure relief piece enables quantitative safety pressure relief during fully enclosed digestion.
Combining nearly 30 years of industry experience, Shanghai Xinyi adopts a safety pressure relief piece structural design to ensure that the digestion tank is completely sealed without leakage under normal working conditions. In an overpressure state, the pressure relief piece automatically lifts to complete the pressure relief, ensuring safe operation. Under normal operation, the safety pressure relief piece will not break and does not need to be replaced, there are no consumables, and there is no risk of leakage.

High-strength aerospace composite fiber outer tank, balancing mechanical strength and anti-corrosion capabilities
High-strength aerospace composite fiber outer tankedge protection). The compressive strength can reach 70MPa (approximately 10,000 psi) and the temperature resistance is 600°C, ensuring the user's operational safety.

Uni-Turn technology synchronous rotation technology improves instrument reliability
The use of Uni-Turn technology enables the upper temperature and pressure sensor and the rotor to rotate synchronously, which ensures high-precision temperature and pressure measurement while avoiding sensor winding, effectively reducing the motor load and improving instrument reliability.

Fully sealed high-pressure digestion technology effectively improves recovery rate and ensures the accuracy of data results.
Ultra-high pressure TFM digestion tank, safety pressure relief piece is responsible for pressure bearing and over-pressure safe release, skirt seal, fully sealed high-pressure digestion mode will not release pressure during the entire digestion process, and the digestion capacity is improved. It has obvious advantages for difficult-to-digest samples and samples that are prone to high pressure, especially for elements that are prone to volatile losses. The test results are satisfactory.

Excellent digestion capacity to meet the digestion needs of special samples
1800W large microwave output power, variable frequency non-pulse continuous microwave emission, and dual magnetrons arranged in two dimensions and two directions, making the microwave field more uniform and energy utilization higher, meeting the digestion needs of special samples.

The instrument and software are easy and fast to operate, and are equipped with a variety of digestion solutions.
The instrument is equipped with a rich and safely verified digestion method library, which is compatible with standard methods such as US-EPA and ASTM. At the same time, users can also edit and store their own method library as needed. The high-definition LED display can display experimental parameters and curve changes in real time, with a friendly interface and easy operation.
One machine has multiple uses. One host can perform microwave digestion/extraction/synthesis, etc.
MDS-15 can carry out microwave digestion, extraction, synthesis and other tasks. It is a multi-purpose machine, giving full play to the characteristics and advantages of microwaves, providing more convenience and possibilities for scientific and technological workers and engineers from all walks of life. Due to the ultra-high working pressure of MDS-15, it can completely digest some special samples, such as petrochemical samples, polymer materials, animal and vegetable oils, etc.; it is also suitable for some high-temperature and high-pressure synthesis systems.
| MDS-15 high-throughput sealed microwave digestion/extraction/synthesis workstation parameters: | |
| power supply | 220-240 VAC 50/60Hz 16A |
| microwave source | 2450MHz, dual magnetron high-energy microwave field emission |
| Installed power of the whole machine | 3600W |
| Microwave output power | 1800W |
| Microwave oven cavity | 65L large volume 316L stainless steel chamber, multi-layer corrosion-resistant PFA spray coating inside and outside |
| Safety explosion-proof door design | Six-layer steel structure self-pop-up safe and explosion-proof sliding furnace door, electronic/mechanical dual control |
| Pressure measurement control system | Piezoelectric crystal pressure sensor, pressure control range: 0-15MPa (about 2200psi), accuracy ±0.01MPa |
| Temperature measurement control system | High-precision platinum resistance temperature sensing, temperature measurement range: 0-350℃, temperature control Accuracy ±0.1℃. Optional IR full tank scanning temperature measurement system, temperature measurement range: 0~350℃ |
| passive protection system | It adopts the top safety bolt design to release overpressure in a directional and quantitative manner. |
| Furnace cavity exhaust system | Cooling time to room temperature is less than 15 minutes |
| Working environment temperature/humidity | 0-40 °C / 15-80% RH |
| Physical dimensions of the whole machine Net weight | 580*690*700mm (width x depth x height) ,72 KG |
| Rotor type | independent framework | Alloy turntable integrated frame |
| Reaction tank model | GP-100 | XXP-100 |
| Batch size | Standard configuration: 15 positions, can be expanded to 16 positions | 18 |
| Inner tank material | TFM | TFM |
| Outer tank material | Aerospace composite fiber | Aerospace composite fiber |
| Reaction tank volume | 100mL | 100mL |
| design temperature | 310℃ | 320℃ |
| design pressure | 15MPa(2200psi) | 18MPa (about 2600psi) |
| picture | ![]() | ![]() |