Fully automatic infrared spectrophotometer

PRICE
Negotiable
Model:OIL510D
Brand:Spectra Invent/Chinainvent
Origin:China

Core Specifications

Instrument type
laboratory
Instrument principle
Infrared spectrophotometry
Measuring range
0~64000mg/l
detection limit
0.06mg/L
Accuracy
Error ≤ 1% (30mg/L~50mg/L)
resolution
0.0005mg/L
Applications
Environmental & WaterChemicals & PetrochemicalsMachinery & EquipmentEnergy & UtilitiesMaterials & Metals

Product Details


OIL510® Series Fully Automatic Infrared Spectrophotometer

Application areas:It can be used to detect the content of oil, petroleum or animal and vegetable oils in industrial wastewater, domestic sewage, oil mist, soil, solid waste and other materials.

Included models:OIL510®B, OIL510®D, OIL510®E, OIL510®F

Meet standards

HJ 637-2018, HJ 1051-2019, HJ 1077-2019, GB/T 5750.7-2023, GB/T 12152-2007, GB 5084-2021, GB 5085.6-2007, etc.

Technical principles

According to the HJ 637-2018 standard, the water sample is extracted with tetrachlorethylene under the condition of pH ≤ 2, and then the oil is measured; the extract is adsorbed with magnesium silicate to remove polar substances such as animal and vegetable oils, and then the petroleum is measured. The content of oil and petroleum is determined by the wave numbers of 2930cm⁻¹ (stretching vibration of C-H bond in CH2 group), 2960cm⁻¹ (stretching vibration of C-H bond in CH3 group) and 3030cm⁻¹ (stretching vibration of C-H bond in aromatic ring). The absorbance A2930, A2960 and A3030 at are calculated according to the correction coefficient; the content of animal and vegetable oils is the difference between oil and petroleum content.

System composition

  • Pre-processing unit:Automatically complete volume reading, quantitative reagent filling, extraction, oil-water separation, magnesium silicate adsorption, pipeline cleaning, and waste liquid collection
  • Detection unit:Automatically scan and detect samples processed by the pre-processing unit, and can also quickly switch to manual mode
  • Data processing system:Integrate information entry, scanning collection, calculation and analysis; meet carbon tetrachloride/tetrachlorethylene dual mode
  • Waste liquid treatment unit:Equipped with waste gas and liquid treatment and collection device, reagents and water are completely separated

Features

✔ Comply with national standards
✔Long life of light source
✔ Automatic baseline adjustment
✔ Non-decentralized functionality
✔ Carbon tetrachloride/tetrachlorethylene
✔ Coefficient/curve dual calibration
✔ Reagent acceptance function
✔ Perfect after-sales service
✔ Automatic monitoring and early warning
✔ Automatic dilution
✔ Extraction rate ≥95%
✔ High degree of automation

Technical parameters

Parameter items Technical indicators
Wavenumber range 3400cm⁻¹~2400cm⁻¹ (i.e. 2941nm~4167nm)
Absorbance range 0.0000~2.0000AU (i.e. transmittance 1~100%T)
Wavelength accuracy ±1nm (i.e. wavenumber repeatability ±0.5cm⁻¹)
Measuring range 0 ~ 64000mg/L, automatic dilution over range
Instrument detection limit 0.06mg/L
linear correlation coefficient r>0.9995
accuracy error ≤1%
Repeatability RSD≤0.6% (30~40mg/L oil standard sample measured 11 times)
Correction coefficient accuracy ±10%
Sampling volume 0~650mL / 0~1000mL (water sample cup with scale)
Analysis time 15 minutes for a single sample, 10 minutes per sample for subsequent samples during continuous multi-sample measurement
Power supply 220V 50Hz 50W
Usage environment -5 ~ 45℃ 20% ~ 85%