Fully automatic infrared spectrophotometer
Core Specifications
Product Details


OIL510® Series Fully Automatic Infrared Spectrophotometer
Application areas:It can be used to detect the content of oil, petroleum or animal and vegetable oils in industrial wastewater, domestic sewage, oil mist, soil, solid waste and other materials.
Included models:OIL510®B, OIL510®D, OIL510®E, OIL510®F
Meet standards
HJ 637-2018, HJ 1051-2019, HJ 1077-2019, GB/T 5750.7-2023, GB/T 12152-2007, GB 5084-2021, GB 5085.6-2007, etc.
Technical principles
According to the HJ 637-2018 standard, the water sample is extracted with tetrachlorethylene under the condition of pH ≤ 2, and then the oil is measured; the extract is adsorbed with magnesium silicate to remove polar substances such as animal and vegetable oils, and then the petroleum is measured. The content of oil and petroleum is determined by the wave numbers of 2930cm⁻¹ (stretching vibration of C-H bond in CH2 group), 2960cm⁻¹ (stretching vibration of C-H bond in CH3 group) and 3030cm⁻¹ (stretching vibration of C-H bond in aromatic ring). The absorbance A2930, A2960 and A3030 at are calculated according to the correction coefficient; the content of animal and vegetable oils is the difference between oil and petroleum content.
System composition
- Pre-processing unit:Automatically complete volume reading, quantitative reagent filling, extraction, oil-water separation, magnesium silicate adsorption, pipeline cleaning, and waste liquid collection
- Detection unit:Automatically scan and detect samples processed by the pre-processing unit, and can also quickly switch to manual mode
- Data processing system:Integrate information entry, scanning collection, calculation and analysis; meet carbon tetrachloride/tetrachlorethylene dual mode
- Waste liquid treatment unit:Equipped with waste gas and liquid treatment and collection device, reagents and water are completely separated
Features
Technical parameters
| Parameter items | Technical indicators |
|---|---|
| Wavenumber range | 3400cm⁻¹~2400cm⁻¹ (i.e. 2941nm~4167nm) |
| Absorbance range | 0.0000~2.0000AU (i.e. transmittance 1~100%T) |
| Wavelength accuracy | ±1nm (i.e. wavenumber repeatability ±0.5cm⁻¹) |
| Measuring range | 0 ~ 64000mg/L, automatic dilution over range |
| Instrument detection limit | 0.06mg/L |
| linear correlation coefficient | r>0.9995 |
| accuracy error | ≤1% |
| Repeatability | RSD≤0.6% (30~40mg/L oil standard sample measured 11 times) |
| Correction coefficient accuracy | ±10% |
| Sampling volume | 0~650mL / 0~1000mL (water sample cup with scale) |
| Analysis time | 15 minutes for a single sample, 10 minutes per sample for subsequent samples during continuous multi-sample measurement |
| Power supply | 220V 50Hz 50W |
| Usage environment | -5 ~ 45℃ 20% ~ 85% |