After the "Formaldehyde Cabbage" incident, Wufeng Instruments' HPLC solution makes formaldehyde nowhere to hide.

Instrument.com.cn Oct 13, 2026

In August 2026, a video of "cabbage dipped in formaldehyde" to preserve freshness quickly fermented on the short video platform, triggering strong public concern about the illegal addition of formaldehyde to vegetables and water products. Entering the national "Quality Month" in September, special sampling inspections of formaldehyde in food by local market regulatory authorities are still continuing, and the demand for formaldehyde testing solutions from food production companies and third-party testing agencies has suddenly increased.

In fact, formaldehyde testing is not without standards. GB 5009.307-2025 "National Food Safety Standard - Determination of Formaldehyde in Food" has been officially implemented in 2025, expanding the testing scope from a single alcoholic beverage to all categories of food.However, one year after the standard was implemented, many laboratories are still facing the same problem: formaldehyde has a small molecule, strong polarity, and no UV absorption. How can it be measured by conventional liquid chromatography?


Please keep this set of solutions from Wufeng Instruments!



Wu

rich

●

square

case

picture

01

Solution core


Derivatization+HPLC-UV


In response to this pain point, our Wufeng Instruments established a liquid chromatography detection method for formaldehyde in food based on the LC100 liquid chromatograph (binary pump + UV-visible detector) and with reference to GB 5009.307-2025.

Key idea: Formaldehyde itself does notUV chromophore, direct detection sensitivity is extremely low. This solution uses pre-column derivatization - formaldehyde reacts with 2,4-dinitrophenylhydrazine (DNPH) under acidic conditions to generate stable formaldehyde-DNPH hydrazone derivatives. This derivative has strong UV absorption at 350nm and is perfectly suitable for UV detectors.


02

Chromatographic conditions


• Column:C18 column(4.6×250mm,5μm)

• Mobile phase: acetonitrile-water (50:50) isocratic elution

• Column temperature: 30℃

• Injection volume: 20μL

• Detection wavelength: 350nm


The data speaks for itself: sensitivity, repeatability, and linearity fully meet standards;

Spectral performance: Regardless of whether the formaldehyde-DNPH derivative standard solution is entered at a low concentration (0.25 mg/L) or a high concentration (4 mg/L), the target peaks have good symmetry, sharp peak shapes, and no tailing.


Sensitivity - far beyond the national standard requirements:

• Input 0.0125mg/L standard solution, the signal-to-noise ratio SN≥3 is determined as the detection limit.

• Add 0.025mg/L standard solution, the signal-to-noise ratio SN ≥ 10, determine the limit of quantification.

• This means that the sensitivity of LC100 is significantly better than the requirements for formaldehyde detection in the national standard, and can easily meet the quantitative requirements for trace amounts of formaldehyde.

Repeatability - RSD as low as 0.054%:

Continuously inject 6 needles of low concentration (0.25mg/L) standard solution, retention timeRSDThe peak area RSD is 0.227%. The instrument has excellent repeatability and fully meets the stability requirements for daily testing.

Linear——R²>0.9999:

Formaldehyde-DNPH derivatives have good linearity in the range of 0~4 mg/L, the linear equation y = 740.79978x + 0.93924, the correlation coefficient R² = 0.9999863, and the quantification is accurate and reliable.


picture

Data speaks, spectral representation

picture
picture



three

Heavy

●

price

value

picture

Repeatability - RSD as low as 0.054%:

Six injections of low-concentration (0.25 mg/L) standard solution were continuously injected. The retention time RSD was 0.054% and the peak area RSD was 0.227%. The instrument has excellent repeatability and fully meets the stability requirements for daily testing.

Linear——R²>0.9999:

Formaldehyde-DNPH derivatives have good linearity in the range of 0~4 mg/L, the linear equation y = 740.79978x + 0.93924, the correlation coefficient R² = 0.9999863, and the quantification is accurate and reliable.


Actual sample verification:

picture

(Chinese cabbage)

picture

(beer)

The method not only stays at the standard level, but is also verified in real food matrices:

• Cabbage: After pre-treatment, no formaldehyde was detected.

• Beer: The detected formaldehyde content was 0.036mg/L, which is far lower than the formaldehyde limit for beer (≤2.0mg/L) in GB 2758-2012 "National Food Safety Standard Fermented Wine and its Prepared Wine", and also lower than the formaldehyde limit (0.9mg/L) in GB 5749-2022 "Hygienic Standard for Drinking Water". It is a natural background substance in the beer brewing process.

This shows that this method can work stably in complex matrices such as vegetables and wine, and can effectively distinguish between "natural background" and "artificial additions".

picture

Solution value

This solution has threefold value:


1. Compliance: Fully refer to GB 5009.307-2025, covering formaldehyde testing of all categories of food, and meeting compliance requirements for regulatory sampling and corporate self-inspection.

2. Sensitivity advantage: The detection limit is as low as 0.0125mg/L, and the quantitation limit is 0.025mg/L. The sensitivity is significantly better than the national standard requirements, providing reliable guarantee for trace formaldehyde detection.

3. Credibility of domestic instruments: The LC100 liquid chromatograph (binary pump + UV) demonstrated excellent peak shape, repeatability and linearity in this solution, providing solid data support for the application of domestic instruments in the field of food testing.


This protocol is only for reference for method development and instrument performance verification. The data in this article are based on laboratory spiked samples and do not constitute a test conclusion for any actual sample or batch.