Determination of iron content in boiler water and cooling water
Boiler water and cooling watermiddleironElement contentDetermination of
Preface
In modern industrial production, the stable operation of boilers and cooling water systems directly affects equipment safety, energy efficiency and lifespan. Monitoring iron levels in water is a key step in assessing water quality and preventing corrosion and scaling. Iron ions may come from pipe corrosion, water contamination orCraftsmanshipImproper handling, overdoseaccumulationIt will reduce heat transfer efficiency, induce local corrosion and accelerate equipment wear. Therefore, accurate determination of iron content is of great significance to ensure system safety and optimize water quality management.This study is based on the standard "GB/T 14427-2017 Boiler water and cooling water analysis methods Determination of iron》,PickuseA500Zeeman atomic absorption spectrometer, taking advantage of its excellent background suppression ability and excellent stability,rightBoiler water and cooling watermiddleironElement contentPerform method validation.
Experimental conditions
• Instrument reagents
instrument:A500Zeemangraphite furnaceatomic absorption spectrophotometer
Reagents:Nitreacid
standard material:GBW(E)084937 Iron single element solution standard material
• Instrument parameters
Instrumental analysis conditions for graphite furnace method
wavelength(nm) | 248.33 |
Spectral bandwidth (nm) | 0.1 |
Negative high pressure (V) | 427.75 |
Element lamp current (mA) | 4.0 |
Button background method | Zeeman buckle background |
Sample injection volume (μL) | 30 |
Temperature rising program of graphite furnace method
step | temperature(℃) | Heating time (S) | hold time (S) | Internal air flow |
1 | 120 | 10 | 30 | big |
2 | 150 | 10 | 5 | big |
3 | 2300 | 0 | 3 | close |
4 | 2400 | 0 | 2 | big |
Data results
• standard curve
• Quantification limit
Number of measurements | 1 | 2 | 3 | 4 | 5 | 6 |
Absorbance | 0.00283 | 0.00237 | 0.00278 | 0.00066 | 0.00368 | 0.00008 |
Number of measurements | 7 | 8 | 9 | 10 | 11 | |
Absorbance | 0.00164 | 0.00237 | 0.00223 | 0.00246 | 0.00442 | |
Standard deviation S of absorbance measured 11 times0 | 0.001117 | |||||
standard curve equation | A=0.042586C-0.002506 | |||||
Correlation coefficient | 0.99979 | |||||
Limit of quantitation (μg/L) | 0.262 | |||||
• Precision
Number of measurements | 1 | 2 | 3 | 4 |
Absorbance | 0.15599 | 0.15658 | 0.15922 | 0.15839 |
Number of measurements | 5 | 6 | 7 | |
Absorbance | 0.15835 | 0.15892 | 0.16059 | |
Absorbance average | 0.15829 | |||
RSD/% | 0.97 | |||
in conclusion
based onA500Zeeman atomic absorption spectrophotometersuperioranalyze performance,This studyEstablisheda measureBoiler water and cooling watermiddleironElement contentreliablemethod. The instrument uses Zeeman background correction technology and real-time dual-beam optical path design, which significantly improves the stability and accuracy of detection and ensuresironReliable determination of elements.The quantitative limit of this method can reach0.262μg/L,Relative standard deviation (RSD)for0.97%,It fully reflects the high sensitivity and good repeatability of the instrument and is suitable for iron content in actual water samples.elementcontentof semenallowanalyze.