Thundermagnetic DDSJ-308F conductivity meter
Core Specifications
Product Details
Thundermagnetic DDSJ-308F conductivity meter
【Main features】
● Adopt color high-definition LCD screen with clear display
● Optional temperature units:℃ and °F;
● Referencetemperature20℃or25℃Optional
● Supports functions such as power-on self-diagnosis, automatic shutdown, power-off protection and factory reset
● Support firmware upgrade function, allowing function expansion and personalized requirements
● supportIP54Protection level
● Comes standard with composite conductivity electrode, electrode holder, dust cover and calibration solution
【Intelligent detection、Automatic identification]
● Intelligent determination of end point,supportAutomatic reading, scheduled reading, scheduled interval reading, manual reading
● Supportautomatic/Manualtemperature compensation
● Automaticidentify4kindGBConductivity standard solution,support1-3point conductivity electrodeCalibration
●Supports multiple conductivity compensation methods such as non-compensation, linear, pure water compensation, etc.
● supportautomaticFrequency switching, matching1.0constant electrodeCoverableCompleteRange measurement(20μS/cmThe following samples are recommended0.1constant or0.01Constant conductivity electrode measurement)
【datamanage,Information traceability]
● Support data storage(each500set), check, delete, transferand print
● Supports multiple search methods, and the search results are displayed in a curved manner
● conform toGLP,Realize data traceability
● haveRS-232interface, supporting connection standardsRS-232serial portPrinter to print measurement results directly,Print format optional
● HaveUSBinterface, through dedicated communication software withPCConnect to realize data transmission

Thundermagnetic DDSJ-308F conductivity meter
【Technical parameters】
model Technical parameters | Thundermagnetic DDSJ-308F conductivity meter | |
instrument level | 0.5class | |
Measurement parameters | Conductivity, resistivity, total dissolved solids (TDS), salinity value, temperature value | |
Conductivity | scope | 0.000μS/cm~1000mS/cm |
minimum resolution | 0.001μS/cm, according to the rangeAutomatic switching | |
Electronic unit reference error | ±0.5% FS | |
Resistivity | scope | 5.00Ω.cm~100.0MΩ.cm |
minimum resolution | 0.01Ω·cm, according to the rangeAutomatic switching | |
Electronic unit reference error | ±0.5% FS | |
TDS | scope | 0.000 mg/L~1000g/L |
minimum resolution | 0.001mg/L, according to the rangeAutomatic switching | |
Electronic unit reference error | ±0.5% FS | |
salinity | scope | (0.00~8.00)% |
minimum resolution | 0.01% | |
Electronic unit reference error | ±0.1% | |
temperature | scope | (-10.0~135.0)℃/(14.0-275.0)℉ |
minimum resolution | 0.1 ℃/0.1℉ | |
Electronic unit indication error | ±0.1 ℃ | |
power supply | Power adapter (input:AC100~240V, output:DC9V) | |
Dimensions (mm),weight(kg) | 242×195×68,0.9 | |