Liquid chromatography mass spectrometry-Anyipu TQ9100 method for detecting the content of fat-soluble vitamins in human serum
This article uses the Anyeep TQ9100 ultra-high performance liquid chromatography tandem mass spectrometry system to establish a method for the simultaneous detection of five fat-soluble vitamins in human serum. The method has good linearity and the quantitative limit meets the requirements of relevant standards. The method and equipment can be used to detect and quantify fat-soluble vitamins.
dimensionVitamin deficiency is mostly a progressive disease with no obvious clinical manifestations. However, long-term subclinical deficiency or deficiency will produce characteristic clinical manifestations. For example, vitamin A deficiency can easily lead to night blindness, vitamin D deficiency can affect bone development, vitamin E deficiency can easily cause reproductive dysfunction, and vitamin K deficiency can affect coagulation function. Clinically carried out systematic vitamin testing projects can accuratelylandReflects the human body's vitamin needs, guides vitamin nutritional supplementation, and is especially suitable for nutritional monitoring of specific groups of people.
This method uses the Anyeep TQ9100 liquid chromatography tandem mass spectrometry system to detect vitamin A (VA), 25-hydroxyvitamin D2 (25OHVD2), 25-hydroxyvitamin D3 (25OHVD3), vitamin E (VE) and vitamin K1 (VK1) in human serum. It supports ESI source and APCI source at the same time, and can be detected without switching ionization sources or changing chromatographic conditions.
Instruments and reagents
Anyeep TQ9100 ultra-high performance liquid chromatography tandem mass spectrometry system
Methanol (mass spectrometry pure, Thermo Fisher Company, USA)
Water (ultrapure water)
Formic acid (Merck, Germany)
liquid phase conditions
mobile phaseA: 0.1% formic acid, water
Mobile phase B: 0.1% formic acid, methanol
column:Phenomenex Kinetex F5 Column 3.0 mm×50 mm 2.6 μm
Column temperature:40 ℃
Flow rate: 0.3 mL/min
Injection volume:10 μL
Mass spectrometry conditions
ESI
Ionization source: Electrospray ionization source (ESI+)
source temperature: 500 ℃
Spray voltage: 5000 V
curtain air: 20 psi
atomizing gas: 50 psi
Auxiliary gas: 55 psi
collision gas: 2
Scan mode:MRM
APCI
Ionization source: Electrospray ionization source (APCI+)
source temperature: 500 ℃
Spray voltage: 5000 V
curtain air: 20 psi
atomizing gas: 50 psi
collision gas: 2
Scan mode:MRM
Gradient elution procedure(ESI)
timemin | A% | B% | flow ratemL/min |
0 | 30 | 70 | 0.4 |
0.5 | 20 | 80 | 0.4 |
2.5 | 10 | 90 | 0.4 |
2.6 | 0 | 100 | 0.4 |
5 | 0 | 100 | 0.4 |
5.1 | 30 | 70 | 0.4 |
6.0 | 30 | 70 | 0.4 |
Gradient elution procedure(APCI)
timemin | A% | B% | flow ratemL/min |
0 | 50 | 50 | 0.6 |
2.5 | 10 | 90 | 0.6 |
3.5 | 0 | 100 | 0.6 |
4 | 0 | 100 | 0.6 |
4.1 | 50 | 50 | 0.6 |
5.5 | 50 | 50 | 0.6 |
MRM parameters of the analyte
serial number | Object to be tested | Q1 ion | Q3 ion | Dwell time (ms) | CE (V) | DP (V) |
1 | VA | 269.1 | 93.1* | 20 | 50 | 50 |
VA | 269.1 | 119.3 | 20 | 15 | 50 | |
internal standard | VA-d6 | 275.1 | 96.3 | 20 | 35 | 75 |
2 | 25OHVD2a | 413.2 | 355.3* | 20 | 2 | 85 |
25OHVD2b | 395.5 | 271.3* | 20 | 25 | 85 | |
internal standard | 25OHVD2-d3a | 416.2 | 358.2* | 20 | 2 | 85 |
25OHVD2-d3b | 398.5 | 274.3* | 20 | 25 | 85 | |
3 | 25OHVD3a | 401.2 | 365.4* | 20 | 2 | 100 |
25OHVD3b | 383.3 | 365.3* | 20 | 5 | 100 | |
internal standard | 25OHVD3-d6a | 407.2 | 368.3* | 20 | 2 | 100 |
25OHVD3-d6b | 389.2 | 368.3* | 20 | 5 | 100 | |
4 | VE | 431.4 | 165.1* | 20 | 50 | 50 |
VE | 431.4 | 137.2 | 20 | 20 | 50 | |
internal standard | VE-d6 | 437.4 | 171.2 | 20 | 25 | 90 |
5 | VK1 | 451.4 | 187.1* | 20 | 20 | 110 |
VK1 | 451.4 | 105.1 | 20 | 50 | 110 | |
internal standard | VK1-d4 | 455.4 | 191.1 | 20 | 20 | 110 |
Note: * is the quantitative ion transition, a is the ESI quantitative ion transition, and b is the APCI quantitative ion transition.
Sample processing
Take 200 samples to be tested μL into a 1.5 mL EP tube, add internal standard solution 20 μL, then add 400 isopropyl alcohol μL; vortex to mix (2500 rpm,5 min) and centrifuged (4 ℃, 14000 rpm,5 min); take the supernatant 500 μL in96In the well plate, blow nitrogen to dryness at 40°C; add 100 μL complex solution (50% acetonitrile), placed on a microplate shaker for 2000 rpm/min shake 2 min,4000 centrifuge for 5 minutes,treatInject.
linear range
serial number | ESIObject to be tested | retention time min | linear range ng/mL | linear equation 1/x2Weighted | Correlation coefficient r2 | Precisiona | Accuracyb |
1 | VA | 2.843 | 20-2000 | Y =0.002006x+0.0450443 | 0.9997 | 7.1% 6.5% | 102.9% 99.2% |
2 | 25OHVD2 | 2.760 | 0.5-50 | Y = 0.014762x+0.020829 | 0.9984 | 8.8% 4.9% | 95.2% 96.0% |
3 | 25OHVD3 | 2.718 | 1-100 | Y= 0.013547x+0.033252 | 0.9994 | 5.5% 3.9% | 99.5% 95.0% |
4 | VE | 4.110 | 200-20000 | Y= 0.000291x+0.021601 | 0.9992 | 5.4% 6.1% | 99.3% 96.7% |
5 | VK1 | 4.152 | 0.1-5 | Y=0.156314x+0.008712 | 0.9968 | 17.9% 7.4% | 94.2% 97.0% |
serial number | APCIObject to be tested | retention time min | linear range ng/mL | linear equation 1/x2Weighted | Correlation coefficient r2 | Precisiona | Accuracyb |
1 | VA | 2.449 | 20-2000 | Y=0.002166x+0.035214 | 0.9979 | 4.2% 5.5% | 99.2% 103.1% |
2 | 25OHVD2 | 2.469 | 1-50 | Y=0.019129x+0.035112 | 0.9986 | 5.2% 7.4% | 105.1% 100.0% |
3 | 25OHVD3 | 2.379 | 2-100 | Y=0.014289x+0.011085 | 0.9996 | 4.1% 6.9% | 105.6% 102.1% |
4 | VE | 4.258 | 200-20000 | Y=0.001055x-0.044219 | 0.9911 | 9.1% 5.9% | 101.1% 90.5% |
5 | VK1 | 4.459 | 0.1-5 | Y=0.274144x+0.008196 | 0.9929 | 6.6% 5.9% | 97.3% 98.9% |
Note: The upstream data of a and b are low concentrations, and the downstream data are high concentrations. The low concentration is 3 times the lower limit of the linear range, and the high concentration is 70% of the upper limit of the linear range.
Summarize
This method uses the TQ 9100 triple quadrupole tandem mass spectrometer system independently developed and produced by Anyipuquan to achieve the detection of fat-soluble vitamins in serum.A、D、E、KSimultaneous detection of。