How to more accurately detect sulfur dioxide residues in food additives and wolfberry? Interpretation of measured data of Heguan HGK-82-6
Let’s talk about the conclusion first: for complex matrix samples, it is recommended todistillationThe time is extended from the normal 15-30 minutes to 90 minutes, and the absorption solution 3% hydrogen peroxide is added to 50 mL. The method is based onGB 5009.34-2022"National Food Safety Standards in Food"sulfur dioxide"Determination", instruments are availableSix-hole fully automatic machine(Our laboratory uses Shanghai Heguan HGK-82-6).
1. Why regular time may not be enough
The determination of sulfur dioxide in food generally follows the distillation-titration route. The national standard for distillation time gives a range, which gives the laboratory flexibility, but complex matrices do not buy it:Compound food additivesHigh in salt and sugar, it is easy to explode when distilled; the dry wolfberry preparation contains pigment and protein.sulfiteIt exists in a bound state and cannot be released at all if there is not enough time; in addition, aldehydes, ketones, etc.volatile componentsInterfering with endpoint judgment, data drift is common. This is also the real source of the "poor reproducibility of the same sample" reported by many laboratories.
2. Actual measurement data: extended to after 90 minutes
We conducted actual testing on samples sent by customers (2026-08-18 report). Reagent system: 0.3 mol/L hydrochloric acid 200 mL + 3% hydrogen peroxide 50 mL + 0.1 mol/LSodium hydroxide titrant, distilled for 90 minutes.
| sample | parallel | Weighing sample quantity (g) | Deduct blank consumption (mL) | SO2 content (%) | Distillation time |
| Compound food additives | 3 parallels | 0.5100 | 10.4 / 10.3 / 11.8 | 6.53 / 6.65 / 7.26 | Average 6.81% |
| Lycium barbarum desiccant | 2 parallels | 2.0041 / 2.0092 | 14.1 / 13.9 | 2.25 / 2.21 | Average 2.23% |
The compound additives were extremely poor at 0.73%, and the wolfberry desiccant was only 0.04% different between the two preparations. Samples with high sugar content and pigments, such as wolfberry, will still be 5%-8% low after 60 minutes, and will be stable after 90 minutes. It is recommended that each laboratory verify this conclusion.
3. Three things are recommended for methodological verification:
If your laboratory is going to use this solution, don’t rush to use it. Do three things: first, parallelism. The same sample should be paralleled more than three times, and the relative range should be controlled within 5%. Second, recovery rate. The recovery rate after adding the standard should be 90%-110%. Third, use third-party certified reference materials or participate in inter-laboratory comparisons to confirm that the data are comparable.
4. Applicable sample range
In addition to compound food additives and wolfberry desiccant, it is also suitable for preserved fruits, wine and fruit wine, dried edible mushrooms, traditional Chinese medicine pieces, spices, soy sauce vinegar, and vermicelli vermicelli. The method complies with both GB 5009.34-2022 and 2020 editions"Chinese Pharmacopoeia", both food and drug supervision scenarios can be covered.
Statement of interests: The measured data in this article were provided by the laboratory of Shanghai Heguan Instrument Co., Ltd. (the instrument in this article is the company's HGK-82-6 model). This answer is for methodological exchange only and does not constitute a purchase recommendation; the specific distillation time should be determined after verification with your own sample matrix.