Seven Key Methods for Ultrapure Water Quality Control in ICP-MS Analysis
I used a high-resolution instrument and selected high-purity reagents, but the results still show deviations?
a lot ofICP-MSOperators have all encountered this confusion: there is no problem with the instrument, the reagents are fine, and the method is fine, but the data is unstable, the blank value is high, and the detection limit cannot be lowered.……
The problem is likely to lie in a link that is easily overlooked——Ultrapure water。
Ultrapure water isICP-MSThe most basic and most used in analysis"Reagents". If the water quality is not up to standard, no matter how expensive the instrument is, it is difficult to guarantee the accuracy of the analysis results.
Below from 7dimensionsHelp you optimize your systemICP-MSexperimental water,
01 Select the right purification column: special ICP type + low boron type
To reach AsiapptevenppqThe detection limit of metal elements is higher than that of conventional purification columns, and ordinary purification columns may not be able to meet the requirements.
Need to adoptSpecial exchange resin for specifically removing metal ions, and accurately mix the proportions of different resins according to special formulas to ensure that the effluent truly meets the trace analysis requirements.
✅Keywords: dedicatedICPtype purification column, low boron type purification column
02 Water inflow must be stable: EDI technology is the key
Ultrapure water"foundation"Depends on incoming water quality.
EDI(Continuous Electrodeionization) ModuleAs the incoming water source of ultrapure water, it not only can efficiently remove metal ions and charged organic matter, but also has excellent removal capabilities for weak electrolytes such as boron and silicon, providing stable and reliable guarantee for back-end ultrapurification.
✅Keywords:EDIWater inflow, stable water supply
03 Be particular about getting water: use it immediately and discard the previous water.
Ultrapure water is not"Keep in reserve"Yes, butTake now and use nowof.
suggestion:
Use freshly prepared ultrapure water;
Abandon the previous agreement10Lwater production, and then collect the subsequent effluent;
Avoid using it directly after standing for a long time.
✅Keywords: Ready to use, discard the water in the front section
04 The environment must be clean: water collection points must also enter a "clean room"
ICP-MSThe host computer is usually placed in a class 1,000 or class 100 clean laboratory, but many people ignore it——Water points are also required。
Airborne particles and aerosols can quickly contaminate ultrapure water. The water terminal should be in a clean environment with the host machine to avoid"last mile"of pollution.
✅Keywords: thousand level/Class 100 laboratory, avoid air exposure
05 Choose the right container material: low dissolution is a hard indicator
If the material of the water container is unqualified, metal ions will directly dissolve and affect the analysis results.
RecommendedPFALow dissolution materialscontainer to prevent the container itself from becoming a source of contamination.
✅Keywords:PFAMaterial, low dissolution
06 Standardized cleaning: three steps and nine vibrations
Cleaning of ultrapure water containers cannot be done carelessly. The following process is recommended:
step | Dosage | frequency | operate |
first step | 60~80 mLUltrapure water | 3Second-rate | every shock10s |
Step 2 | 30~50 mLUltrapure water | 3Second-rate | every shock10s |
Step 3 | 60~80 mLUltrapure water | 3Second-rate | every shock10s |
Place after cleaningClean fume hoodLet dry in medium and set aside.
✅Keywords: standardize cleaning, avoid secondary pollution
07 Regular maintenance: Don’t wait for the alarm to be remembered
Pure water system needsRegular maintenanceTo ensure consistent water quality:
⏰ Replace consumables on time: Refer to the manufacturer’s recommended usage time and water volume;
⚠️ Pay attention to alarm prompts: Maintenance on the display/Don’t ignore maintenance information;
📊 Calibrate monitoring equipment regularly:Including resistivity meter andTOCMonitor to ensure that the data is true and reliable, and corrective measures can be taken in a timely manner when necessary.
✅Keywords: regular maintenance, consumable replacement, equipment calibration
Summarize
From purification column selection, incoming water quality, water intake operation, environmental control, container material, cleaning specifications to system maintenance——Every link may have an impactICP-MSdata quality"invisible killer"。
Put the above7Only by doing the right things can you eliminate the interference of water quality on the analysis results to the greatest extent, making your data cleaner and more credible.