Home/News/Liquid Chromatography-Mass Spectrometry-Method for Detection of 17 Perfluorinated Compounds in Drinking Water
Company News

Liquid Chromatography-Mass Spectrometry-Method for Detection of 17 Perfluorinated Compounds in Drinking Water

Instrument.com.cn Oct 22, 2026

introduction

Perfluorinated compounds are persistent and bioaccumulative,Accumulation levels hundreds to thousands of times higher than known organochlorine pesticides and dioxins。Perfluorinated compounds have various toxicities such as reproductive toxicity, mutagenic toxicity, developmental toxicity, neurotoxicity, and immunotoxicity. They are a type of environmental pollutant with systemic multi-organ toxicity.. "GB 5749-2022 Hygienic Standards for Drinking Water" clearly stipulates the detection limits for perfluorooctanoic acid, perfluorooctane sulfonate, etc. in drinking water.

This article developed a method for the simultaneous detection of 17 perfluorinated compounds in drinking water based on the relevant requirements for the detection of perfluorooctanoic acid in Section 84 of "GB/T 5750.8-2023 Standard Test Methods for Drinking Water Part 8: Organic Indicators".

Instruments and reagents

Anyeep TQ9120Ultra-high performance liquid chromatography tandem mass spectrometry system

Methanol (mass spectrometry pure, Thermo Fisher Company, USA)

Acetic acid, ammonium acetate(Chromatographypure,German Merk Company)

ammoniawater(Analytically pure, Shanghai Anpu)

Mixed standards of 17 perfluorinated compounds(Beijing Tan Mo,81589x)

3 kinds of perfluorinated compounds internal standard mixed standard (Beijing Tanmo, 83216x)

Anion exchange solid phase extraction column(Shanghai Anpu, SBEQ-CA3881)

liquid phase conditions

Mobile phase A:water, containing 5mM ammonium acetate

Mobile phase B: methanol

column:  Waters Acquity UPLC BEH C18,1.7 µm, 2.1 × 100 mm

flow  speed: 0.3 mL/min

column  temperature: 40 ℃

Injection volume: 5 μL

Mass spectrometry conditions

Ionization source: Electrospray ionization source (ESI-)

source temperature Spend: 550 ℃

Spray voltage: 4500 V

gas curtain gas: 30 psi

fog change gas: 60 psi

auxiliary help gas: 60 psi

bump Collide gas: 11

Scan mode:MRM

Gradient elution procedure

time min

A %

B %

Flow rate mL/min

0.0

80

20

0.3

1.5

60

40

0.3

10.0

10

90

0.3

10.1

2

98

0.3

12.0

2

98

0.3

12.1

80

20

0.3

14.0

80

20

0.3

MRM parameters of the analyte

Object to be tested

Q1 ion

Q3 ion

CE(V)

DP(V)

PFBA

perfluorobutyric acid

212

169*

2

30

212

97

2

30

PFPeA

perfluoropentanoic acid

263

219*

45

30

263

69

4

30

PFHxA

perfluorohexanoic acid

313

269*

15

30

313

119

4

30

PFHpA

Perfluoroheptanoic acid

363

319*

14

30

363

169

2

30

PFOA

perfluorooctanoic acid

413

369*

12

30

413

169

2

30

PFNA

Perfluorononanoic acid

463

419*

14

30

463

169

2

30

PFDA

Perfluorodecanoic acid

513

469*

14

30

513

219

5

30

PFUnDA

Perfluoroundedanoic acid

563

519*

15

60

563

319

6

60

PFDoDA

perfluorodecanoic acid

613

569*

20

60

613

169

6

60

PFTrDA

Perfluorotridecanoic acid

663

619*

22

60

663

169

6

60

PFTeDA

perfluorotetradecanoic acid

713

669*

24

90

713

169

10

90

PFHxDA

perfluorohexadecanoic acid

813

769*

28

120

813

169

12

120

PFODA

perfluoroctadecanoic acid

913

869*

38

120

913

169

58

120

PFBS

perfluorobutane sulfonate

299

80*

56

90

299

99

68

90

PFHxS

PFHxS

399

80*

68

120

399

99

70

120

PFOS

PFOS

499

80*

68

150

499

99

80

150

PFDS

perfluorodecanesulfonic acid

599

80*

80

180

599

99

2

180

PFHxA-13C2

315

270

15

30

PFOA-13C4

417

372

12

30

PFOS-13C4

503

80

68

150

Note: * is the quantitative ion transition

Sample processing

Measure 1 LWater sample to be tested,Join 4.625 g ammonium acetate, acetic acidadjustpHto 6.8 - 7.0,Add to each liter of water sample100 μg/LIsotope internal standard mixed standard solution 100 μL,Mix well,0.22 μmcellulose acetate membranefilter.

Use a pre-activated weak anion exchange solid-phase extraction column for extraction (use a mass fraction of 0.1% in sequenceAmmonia-methanol solution5 mL, methanol7 mLand pure water10 mLactivation), control the flow rate at about 8 mL/min, use 5 mL of 25 mM ammonium acetate solution (acetic acid to adjust pH = 4.0) and 12 mL of pure water to rinse, drain for 15 min, and use in sequenceMethanol5 mLand ammonia-methanol solution7 mLperform elution,Collect all eluates。40 ℃Blow nitrogen until dryness, and reconstitute with 1 mL of methanol-water solution (methanol:water 3:7, v/v). 5μL injection.

linear range

Object to be tested

retention time

min

detection limit

μg/L

Method detection limit ng/L

linear range

μg/L

linear equation

1/x2

Correlation coefficient

r2

PFBA

2.31

0.2

0.2

0.5-200

Y = 194.496 * x + 73.7965

0.9935

PFPeA

3.96

0.2

0.2

0.5-200

Y = 1125.76 * x + 15.1260

0.9991

PFBS

4.39

0.01

0.01

0.02-200

Y = 596.905 * x + 1.52550

0.9950

PFHxA

5.43

0.02

0.02

0.1-200

Y = 1473.82 * x + 6.25335

0.9976

PFHpA

6.59

0.02

0.02

0.1-200

Y = 1649.93 * x + 8.86020

0.9978

PFHxS

6.70

<0.01

<0.01

0.01-200

Y = 463.553 * x + 0.59411

0.9952

PFOA

7.48

0.2

0.2

0.5-200

Y = 1920.98 * x + 4.80332

0.9979

PFNA

8.26

0.02

0.02

0.1-200

Y = 2106.58 * x + 23.5707

0.9959

PFOS

8.28

0.01

0.01

0.02-200

Y = 380.290 * x + 1.39007

0.9944

PFDA

8.89

0.02

0.02

0.1-200

Y = 317.343 * x + 1.99813

0.9941

PFDS

9.44

0.01

0.01

0.02-200

Y = 270.518 * x + 0.35341

0.9930

PFUnDA

9.45

0.02

0.02

0.1-200

Y = 2243.58 * x + 15.3498

0.9911

PFDoDA

9.91

0.02

0.02

0.1-200

Y = 109.115 * x + 0.06470

0.9953

PFTrDA

10.31

0.02

0.02

0.1-200

Y = 89.6985 * x + 0.51377

0.9908

PFTeDA

10.64

0.1

0.1

0.2-200

Y = 1547.08 * x + 21.4153

0.9912

PFHxDA

11.19

0.2

0.2

0.5-200

Y = 450.902 * x + 21.1795

0.9979

PFODA

11.61

0.2

0.2

0.5-200

Y = 109.504 * x + 14.6840

0.9935

Summarize
In this paper, an ultra-high performance liquid chromatography tandem mass spectrometry system (Anyee p TQ9120) Analysisdrinking watermiddle17 perfluorinated compoundsdetection method. It can be seen from the results that this method has excellent sensitivity, stability and linear range.The recovery rate meets the national standard requirements.Available forPerfluorinated compounds in drinking waterRoutine analytical testing.