InstrumentCube
Products

Home/News/Method for determination of 36 antibiotic drugs in pork using Anyeep TQ9120 ultra-high performance liquid chromatography tandem mass spectrometry system
Company News

Method for determination of 36 antibiotic drugs in pork using Anyeep TQ9120 ultra-high performance liquid chromatography tandem mass spectrometry system

Instrument.com.cn Oct 23, 2026

introduction:

Tetracyclines, sulfonamides and quinolones are antibacterial veterinary drugs widely used in livestock and poultry breeding. Excessive abuse can easily lead to excessive drug residues in animal foods. Long-term intake of meat, eggs, milk and other foods with excessive residues can cause accumulation of drug resistance, allergic reactions and organ damage in the human body, seriously threatening food safety and human health.

Currently, Class III veterinary drug residue contamination is widespread, covering a variety of livestock and poultry product matrices. Common test samples include animal tissues such as muscle, liver, kidney, and fat of pigs, cattle, sheep, and chickens. The current domestic national food safety standard GB 31658.17-2026 clarifies the testing specifications for multi-component residues of three types of veterinary drugs and adapts to the testing needs of batch samples. The matrix of animal foods is complex and contains interfering substances such as proteins and fats. Most drug residues are at trace levels, and conventional detection methods have insufficient sensitivity and specificity. The establishment of accurate and efficient multi-residue simultaneous detection technology can effectively adapt to national standard testing requirements and provide important technical support for animal food veterinary drug residue screening and food safety risk management and control. This method refers to GB 31658.17-2021 and uses the external standard method for quantification.

Instruments and reagents:

AnyeepTQ9120 ultra high performance liquid chromatography tandem mass spectrometry system

Methanol (mass spectrometry pure, Shanghai Anpu Company)

Formic acid (chromatographically pure, Shanghai Anpu Company)

Ethyl acetate (chromatographically pure, Shanghai Anpu Company)

Mcllvanine-Na2EDTA buffer (analytical grade, Northern Weiye Metrology Group Co., Ltd.)

Sodium dihydrogen phosphate (analytical grade, Aladdin)

Disodium hydrogen phosphate (analytical grade, Aladdin)

Concentrated ammonia (analytical grade, Shanghai Anpu Company)

Liquid phase conditions:

Mobile phase A: 0.1% formic acid-water

Mobile phase B: methanol

Column: Waters ACQUITY UPLC HSS T3 1.8 μm, 2.1*100 mm

Column temperature: 40 ℃

Flow rate: 0.3 mL/min

Injection volume: 5 μL

Mass spectrometry conditions:

Ionization source: electrospray ionization source (ESI+)

Source temperature: 550℃

Spray voltage: 5000V

Curtain air: 45 psi

Atomizing gas: 60 psi

Auxiliary air: 60 psi

Collision gas: 12

Scan mode: MRM

Gradient elution procedure:

time min

A%

B%

Flow rate mL/min

0

90

10

0.3

1

60

40

0.3

3

60

40

0.3

4

5

95

0.3

6

5

95

0.3

6.1

90

10

0.3

9

90

10

0.3

Sample processing:

Weigh 1 g of sample (accurate to ±0.01 g), add 8 mL of Mcllvanine-Na2EDTA buffer, vortex for 1 min, extract with ultrasonic for 20 min, centrifuge at 10000 r/min for 5 min, and collect the supernatant. Add 8 mL of phosphate buffer to the residue, repeat the extraction once, combine the two extractions, mix well, and set aside.

The solid-phase extraction column was activated with 5 mL of methanol and 5 mL of water in sequence. Pass the reserve solution through the column, elute with 5 mL of water and 5 mL of 20% methanol aqueous solution in sequence, drain, and elute with 10 mL of eluent. Collect the eluate, blow dry with nitrogen in a 45°C water bath, add 1 mL of reconstituted solution, vortex for 1 min to dissolve the residue, centrifuge at 14000 r/min for 5 min, filter through a 0.22 μm filter membrane, and take the filtrate for measurement.

Standard curve preparation:

Accurately pipette an appropriate amount of mixed standard working solution and dilute it with blank matrix to 0.1 μg/L, 0.2 μg/L, 0.5 μg/L, 1 μg/L, 2 μg/L, 5 μg/L, 10 μg/L, 20 μg/L, 50 μg/L, 100 μg/L, 200 μg/L, 500 μg/L series standard solutions for liquid chromatography mass spectrometry-mass spectrometry measurement.

Summarize

This paper established a detection method for 4 kinds of tetracyclines, 19 kinds of sulfonamides and 13 kinds of quinolones in pork using the ultra-high performance liquid chromatography tandem mass spectrometry system Anyeep TQ9120. It can be seen from the results that this method has excellent sensitivity, stability and linear range, and can be used for routine analysis and detection of tetracyclines, sulfonamides and quinolones.