InstrumentCube
Products

Huapu Scientific Instrument S6000PLUS high performance liquid chromatograph

PRICE
Negotiable
Model:S6000PLUS 高效液相色谱仪
Brand:Acchrom-Tech
Origin:China

Core Specifications

Instrument type
Conventional HPLC
Application scenarios
General instrument
Applications
Environmental & WaterAgriculture & FoodBiotechnology & Life SciencesConsumer Products & ToysChemicals & PetrochemicalsElectronics & SemiconductorsAutomotive & TransportationTextiles & ApparelPublic Safety & JusticeOther

Product Details

pump system

PLUS pump.png

■ Salt-resistant and acetonitrile-compatible check valve design

Under conditions of high salt and high proportion of acetonitrile, the pressure is more stable to avoid pump pressure loss.

■ Binary/quaternary pump system optional

A binary high-pressure gradient pump and a quaternary low-pressure gradient pump are available, and the pump system can be flexibly configured according to the application scenario.


■The maximum withstand voltage reaches70MPa

The withstand voltage reaches70MPa, fully adaptable2.5umParticle size chromatography column applications.



Sample management system

PLUS injector.png

■Sample residue can reach<0.003%

Innovative inlet design achieves long-life sealing and near-zero residue.

Specially flush the flow path and flushing fluid so that the entire sampling system is fully flushed.

■Temperature-controlled autosampler

The temperature control of the automatic injector uses the Peltier temperature control module to achieve1℃~45Effective temperature control of ℃. The injector temperature can be accurately stabilized at the set value to ensure the stability of temperature-sensitive samples.


■Flow-through needle injection mode, no sample is wasted

The autosampler adopts flow-through needle injection, which solves the problem of wasting sample and causing "needle overflow" in the sample loop injection mode. The sample can be "taken out as much as it is put in".



Column thermostat management system

plus column thermostat.png

■Peltier effect heating/cooling method makes temperature control more stable

Temperature setting range: 1~90℃ (the minimum unit is 1℃).

Can be placed300mmcolumn

■Pipeline preheating function

The mobile phase is preheated before entering the column to reduce interference with the temperature environment inside the column and ensure good chromatographic peak shape symmetry, sharpness and reproducibility.


A wide range of detectors with stable performance

PLUS detector.png

■Various detectors to choose from

Provide complete detection technology, including UV、DAD、FLR、RI、ELSWait for the detector.


Main technical features

Gradient mode:
Binary/quaternary gradient
Flow rate range:
0.001-10.000 ml/min
Maximum withstand voltage:
70MPa
Maximum injection volume:
30ul, expandable to 100ul
Injector temperature:
1-45℃ (24℃ at room temperature-20℃ above room temperature)
Sampler capacity:
120-bit 2ml vial, 96-well plate compatible
Column oven temperature:
1-90℃ (24℃ at room temperature-65℃ at room temperature)
Automatic column switching valve:
Built-in, optional
Maximum column parameters that can be accommodated:
Can accommodate300mm long column
DAD detector
1024 photosensitive devices
Wavelength self-calibration:
Built-in mercury lamp calibration
Maximum sampling rate:
160Hz
Reference background fitting:
Multi-channel independent mode, reference wavelength is defined by channel